Kwaliteit 1898200T M4 Straightshot Microdebrider en Handstuk voor reparatie Fabriek

1898200T M4 Straightshot Microdebrider en Handstuk voor reparatie

1898200T M4 Straightshot Microdebrider and Handpiece for repair ► Brief Description Fault: 1) The bearing is corroded and the seal is broken. 2) Motor stuck, inside is full of disinfection residue, motor corrosion. 3) The rotation direction of the knife head is stuck. 4) Handle paint off, internal ...
Kwaliteit Flexibel endoscoop reparatie training voor Olympus 260 serie Fabriek

Flexibel endoscoop reparatie training voor Olympus 260 serie

Detail Repair Training For Flexible Endoscope Training Detail 1.Learn about the types of OLYMPUS flexible endoscope, the names of their components and processor, etc. 2.Common Faults and Solutions for 260 series endoscopes 3.Detection steps and practice for 290 series endoscopes 4.Disassemble the ...
Kwaliteit SonoScape EG-500 Elektronische Gastroscopie Voor Herstel Fabriek

SonoScape EG-500 Elektronische Gastroscopie Voor Herstel

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The ...
Kwaliteit Olympus ShockPULSE-SE Lithotripsie Handstuk SPL-T Voor Reparatie Fabriek

Olympus ShockPULSE-SE Lithotripsie Handstuk SPL-T Voor Reparatie

Olympus ShockPULSE-SE lithotripsy handpiece SPL-T for repair ► Brief Description Fault: 1.No ultrasonic energy output. Ultrasonic crystal is damaged. Solution: Replace the ultrasonic crystal. ►Problem & Solutoin Brand Olympus Model SPL-T Ultrasonic crystal Damaged Fault No ultrasonic energy output ...
Kwaliteit OLYMPUS OTV-S6H-1D Camera Head Voor reparatie Fabriek

OLYMPUS OTV-S6H-1D Camera Head Voor reparatie

Olympus OTV-S6H-1D Camera Head For Repair ► Brief Description Fault: 1.Image yes. 2.Cable aging and interference Solution: Replace new cable. ►Problem & Solutoin Brand Olympus Model OTV-S6H-1D Fault Cable aging and interference Solution Replace cable ► Applications Laparoscopy (abdominal surgery) ...
Kwaliteit S5-1 sector-array-ultrasone-sonde voor reparatie Fabriek

S5-1 sector-array-ultrasone-sonde voor reparatie

S5-1 sector array ultrasound probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin ModelS5-1FaultArray element aging, lens bubbleSolutionReplace Crystal ► Common issues with S5-1 Probe Image Dropouts / Dead ElementsAcoustic Lens ...
Kwaliteit OLYMPUS CF-H180AL Elektronisch Colonoscoop voor reparatie Fabriek

OLYMPUS CF-H180AL Elektronisch Colonoscoop voor reparatie

OLYMPUS CF-H180AL Electronic Colonoscope For Repair ► Brief Description Inspection Results: 1. The image is dark, and the CCD is aged. 2. One branch of the light beam is disconnected. 3. The insertion tube is aged and wrinkled. 4. The lower angle is malfunctioning, the other angles are tight, and ...
Kwaliteit Olympus TJF-Q180V Elektronisch Duodenoscoop Voor Reparatie Fabriek

Olympus TJF-Q180V Elektronisch Duodenoscoop Voor Reparatie

OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair ► Brief Description Inspection Results: 1. Image display is available; 2. The light bundle brightness is acceptable; 3. The insertion tube is worn; 4. The light guide tube is worn; 5. The angle is insufficient, the snake bone, steel wire and helix ...
Kwaliteit Aloka UST-5413 lineaire array echografie probe voor reparatie Fabriek

Aloka UST-5413 lineaire array echografie probe voor reparatie

Aloka UST-5413 linear array ultrasound probe for repair ► Brief Description Fault: Lens break. Solution: Replace lens. ►Problem & Solutoin BrandAlokaModelUST-5413FaultLens breakSolutionReplace lens ► Common issues with UST-5413 Probe Image dropouts or black linesAcoustic lens damageCable damageConne...
Kwaliteit Ultrasonix EC9-5/10 Endokavitarische sonde voor reparatie Fabriek

Ultrasonix EC9-5/10 Endokavitarische sonde voor reparatie

Ultrasonix EC9-5/10 Endo-cavitary probe for repair ► Brief Description Fault: Bad array element. Solution: Replace crystal and housing. ►Problem & Solutoin BrandUltrasonix ModelEC9-5/10FaultBad array elementSolutionReplace crystal and housing ► Common issues with EC9-5/10 Probe Image Dropouts or ...
Kwaliteit GE 4D3C-L 4D Convex Transducer voor reparatie Fabriek

GE 4D3C-L 4D Convex Transducer voor reparatie

GE 4D3C-L 4D Convex Transducer For Repair ► Brief Description Fault: 1.Probe dome abrasion & leakage, 2. Dome tube damaged, 3. 4D function not work, 4. Electric motor damaged. Solution:Replace the dome, dome tube; repair electric motor. ►Problem & Solutoin BrandGEModel4D3C-LFault1.Probe dome ...
Kwaliteit PENTAX Medical EPK-i7000 Videoprocessor Voor Reparatie Fabriek

PENTAX Medical EPK-i7000 Videoprocessor Voor Reparatie

PENTAX Medical EPK-i7000 Video Processor For Repair ► Brief Description Brand: PENTAX (PENTAX Endoscope Processor) Model: EPK-i7000 Serial Number: AC*** Fault: It cannot recognize the endoscope and reports an error. Solution: Repair the system board. Repair Time: 3-5 working days. Warranty Period: ...
Kwaliteit S5-2 Hartsonde voor reparatie Fabriek

S5-2 Hartsonde voor reparatie

S5-2 Cardiac probe for repair Fault: Lens problem, cable broken, without strain relief. Solution: Replace probe lens, repair cable, replace strain relief. ►Problem & Solutoin Model S5-2 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, repair cable, replace strain ...
Kwaliteit L15-7io Lineaire sonde voor reparatie Fabriek

L15-7io Lineaire sonde voor reparatie

L15-7io Linear probe for repair ► Brief Description Fault: Lens problem. Solution: Replace probe lens. ►Problem & Solutoin Model L15-7io Fault Lens problem Solution Replace probe lens ► Common issues with L15-7io probe Cable and Connector Issues Acoustic Lens Damage Crystal/Element Failure Housing...
Kwaliteit C5-1 convexe sonde voor reparatie Fabriek

C5-1 convexe sonde voor reparatie

C5-1 convex probe for repair ► Brief Description Fault:Lens problem, cable broken, without strain relief. Solution:Replace probe lens, replace a new cable, replace strain relief. ►Problem & Solutoin Model C5-1 Fault Lens problem, cable broken, without strain relief Solution Replace probe lens, ...
Kwaliteit EPlQ/X5-1 Probe 2D beeld abnormaal voor reparatie Fabriek

EPlQ/X5-1 Probe 2D beeld abnormaal voor reparatie

EPlQ/X5-1 Probe For Repair ► Brief Description Fault: 2D image abnormal Solution: Replace scanhead ►Problem & Solutoin Model EPlQ/X5-1 Fault 2D image abnormal Solution Replace scanhead ► Common Problems & Symptoms Image Artifacts (black lines, dropouts, flickering) → Damaged elements or cable No ...
Kwaliteit C5-2 Convex Sonde Voor Reparatie Medische Apparaat Onderdelen Fabriek

C5-2 Convex Sonde Voor Reparatie Medische Apparaat Onderdelen

C5-2 Convex Probe For Repair ► Brief Description Fault: Strain relief damage, cable broken. Solution: Replace strain relief, repair cable. ►Problem & Solutoin Model C5-2 Convex Fault Strain relief damage, cable broken Solution Replace strain relief, repair cable ► Common Issues with C5-2 Probes Dead ...
Kwaliteit L12-5 lineaire array ultrageluidsonde voor reparatie Fabriek

L12-5 lineaire array ultrageluidsonde voor reparatie

L12-5 linear array ultrasound probe for repair ► Brief Description Fault: Crystal damage. Solution: Replace crystal. ►Problem & Solutoin Model L12-5 Fault Crystal damage Solution Replace crystal ► Common issues with L12-5 Probe Image Dropouts / Dead Elements Acoustic Lens Damage Cable Damage ...
Kwaliteit L12-3 lineaire sonde voor reparatie Fabriek

L12-3 lineaire sonde voor reparatie

L12-3 linear Probe for repair ► Brief Description Fault: Array element aging, lens bubble. Solution: Replace Crystal. ►Problem & Solutoin Model L12-3 Fault Array element aging, lens bubble Solution Replace Crystal ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket ...
Kwaliteit L12-3 Ultrasone sonde voor reparatie Fabriek

L12-3 Ultrasone sonde voor reparatie

L12-3 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Model L12-3 Fault Lens problem Solution Replace lens ► Common issues with L12-3 Probe Image Dropout: Shadows Dual Images Cable Jacket Damage Connector Damage Strain Relief ...
Kwaliteit Reparatie van medische hulpmiddelen Cases Stryker 30o 0502-503-030 laproscope voor reparatie Fabriek

Reparatie van medische hulpmiddelen Cases Stryker 30o 0502-503-030 laproscope voor reparatie

Stryker 30º 0502-503-030 laproscope, SN:103**** Fault: Image blur, objective lens loose without glue, inner optical parts broken. Solution: Add glue for objective lens, replace inner optical parts, clean and adjust optical system, seal with eyepiece. Repair cost: $*** Common Issues with the Stryker ...
Kwaliteit Reparatie van medische hulpmiddelen LINVATEC HD 4300 Artroscoop voor reparatie Fabriek

Reparatie van medische hulpmiddelen LINVATEC HD 4300 Artroscoop voor reparatie

LINVATEC HD 4300 Arthroscope for repair, SN: 62**** Fault: Image with black shadow; eyepiece with scratches, inner optical parts broken. Replace objective lens, inner optical parts, clean and adjust optical system, seal with eyepiece. Image Quality Related Issues: 1.Fogging or Blurry Vision: This is ...
Kwaliteit FUJINON EG-530N gastroscoop voor reparatie Fabriek

FUJINON EG-530N gastroscoop voor reparatie

FUJINON EG-530N gastroscope for repair ► Brief Description Fault: 1.Biopsy tube broken; 2.Angle is not enough; 3.Camera over-heating, there are occasional snowflakes on the image and big risk for CCD; 4.Insert tube aging and wrinkle; 5.Light guide tube aging and wrinkle; 6.Stress boot of light guide ...
Kwaliteit Olympus A4674A Autoclaveerbare Hysteroscoop voor reparatie Fabriek

Olympus A4674A Autoclaveerbare Hysteroscoop voor reparatie

Olympus A4674A Autoclavable Hysteroscope for repair ► Brief Description Fault: 1. Image blur and black shadow. 2. Rod lens broken. 3. Outer tube bending and transformation. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the ...