Samsung Medison P2-4AH Hartsonde voor reparatie

Samsung Medison P2-4AH Cardiac probe for repair ► Brief Description Fault:Lens problem, crystal damage. Solution:Replace probe lens, replace crystal. ►Problem & Solutoin Brand Samsung Model P2-4AH Fault Lens problem, crystal damage Solution Replace probe lens, replace crystal ► Common issues with P2-4AH probe Image Dropout (Dead Elements) Cable Damage Connector Damage Acoustic Lens/Membrane Damage If your instrument has any of the above problems or others, please feel free to
Kwaliteit OLYMPUS TJF-Q180V DUODENOSCOOP VOOR REPARATIE Fabriek

OLYMPUS TJF-Q180V DUODENOSCOOP VOOR REPARATIE

OLYMPUS TJF-Q180V DUODENOSCOPE FOR REPAIR ► Brief Description Fault: 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. Solution: Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD). ►Problem & Solutoin Brand OLYMPUS Model TJF-Q180V Fault 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5.
Kwaliteit Olympus CF-240L Colonoscoop voor reparatie Fabriek

Olympus CF-240L Colonoscoop voor reparatie

Olympus CF-240L Colonoscope for repair ► Brief Description Fault: Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch. Solution: Replace rubber, objective lens, C cover. Adjust the angle. ►Problem & Solutoin Brand Olympus Model CF-240L Fault Leakage, glass of light bundle blur, rubber leak, angle not enough, image fuzzle,objective lens scratch, Solution Replace rubber, objective lens, C cover. Adjust the angle. ► Common
Kwaliteit Karl Storz 27005BA endoscoop voor reparatie Fabriek

Karl Storz 27005BA endoscoop voor reparatie

Karl Storz 27005BA endoscope for repair ► Brief Description Fault: 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope and test. ►Problem & Solutoin
Kwaliteit Toshiba PST-25BT Phased Array Probe Voor Reparatie Fabriek

Toshiba PST-25BT Phased Array Probe Voor Reparatie

Toshiba PST-25BT Phased Array Probe For Repair ► Brief Description Fault: Housing break, image weak, crystal aging. Solution: Replace crystal and housing. ►Problem & Solutoin Brand Toshiba Model PST-25BT Fault Housing break, image weak, crystal aging Solution Replace crystal and housing ► Common issues with PST-25BT Phased Array Probe Image Quality Issues Cable and Connector Issues Probe Casing or Acoustic Lens Damage Temperature Abnormalities Host System Cannot Recognize
Kwaliteit STRYKER 1188-210-105 camerakop voor reparatie Fabriek

STRYKER 1188-210-105 camerakop voor reparatie

STRYKER 1188-210-105 camera head for repair ► Brief Description Fault: Image with interrupt. Cable aging. Solution: Replace compatible new cable and connector. ►Problem & Solutoin Brand STRYKER Model 1188-210-105 Fault Image with interrupt. Cable aging Solution Replace compatible new cable and connector ► Common issues with 1188-210-105 camera head Image Quality Problems Physical and Cable Damage Button and Control Malfunctions Overheating If your instrument has any of the
Kwaliteit STORZ Camerakop H3-ZA Voor Reparatie Fabriek

STORZ Camerakop H3-ZA Voor Reparatie

STORZ Camera Head H3-ZA For Repair ► Brief Description Fault:Image streaks frequently,cable loose contact and aged. Solution: Replace cable kit. ►Problem & Solutoin Brand STORZ Model H3-ZA Fault Image streaks frequently,cable loose contact and aged Solution Replace cable kit ► Common issues with H3-ZA Camera Image Quality Problems Connection and Communication Issues Functional Abnormalities Physical Damage If your instrument has any of the above problems or others, please
Kwaliteit GE RIC6-12-D endo-cavitaire probe voor reparatie Fabriek

GE RIC6-12-D endo-cavitaire probe voor reparatie

GE RIC6-12-D endocavity probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens,Clean cable. ►Problem & Solutoin Brand GE Model RIC6-12-D Fault Lens problem Solution Replace lens,Clean cable ► Common issues with RIC6-12-D Probe Dome Damage and Fluid Leaks Image Dropouts (Dead Elements) Cable and Connector Issues Motor/Steering Failure (for 4D functionality) If your instrument has any of the above problems or others, please feel free to contact us for
Kwaliteit Hitachi C41V1 Ultrasone sonde voor reparatie Fabriek

Hitachi C41V1 Ultrasone sonde voor reparatie

Hitachi C41V1 Ultrasuound probe for repair ► Brief Description Fault: Lens problem. Solution: Replace lens. ►Problem & Solutoin Brand Hitachi Model C41V1 Fault Lens problem Solution Replace lens ► Common issues with C41V1 Probe Image Dropouts or Streaks Cable and Connector Damage Acoustic Lens Issues Housing Damage If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you solve the issue. ► About
Kwaliteit Sonoscape 8-L743 lineaire sonde voor reparatie Fabriek

Sonoscape 8-L743 lineaire sonde voor reparatie

Sonoscape 8-L743 linear probe for repair ► Brief Description Fault: Crystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model 8-L743 Fault Crystal problem Solution Replace the crystal ► Common issues with 8-L743 Probe Acoustic Lens Damage Piezoelectric Crystal Damage (Dead Elements) Cable and Connector Issues Cracked Housing If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our
Kwaliteit Sonoscoop C362 Buiksonde voor reparatie Fabriek

Sonoscoop C362 Buiksonde voor reparatie

Sonoscape C362 Abdominal probe for repair ► Brief Description Fault: Cystal problem. Solution: Replace the Crystal. ►Problem & Solutoin Brand Sonoscape Model C362 Fault Cystal problem Solution Replace the crystal ► Common issues with C362 Probe Image Dropouts (Black Lines or Streaks) Damaged Lens Cracked Housing or Assembly Gaps Cable Issues Color Artifacts in the Image If your instrument has any of the above problems or others, please feel free to contact us for assistance.
Kwaliteit GE 4C-RS abdominale probe voor reparatie Fabriek

GE 4C-RS abdominale probe voor reparatie

GE 4C-RS abdominal probe for repair ► Brief Description Fault: Lens broken, cable broken, crystal bad. Solution: Replace probe head, repair cable. ►Problem & Solutoin Bran GE Model 4C-RS Fault Lens broken, cable broken, crystal bad Solution Replace probe head, repair cable ► Common issues with 4C-RS Probe Image dropouts / Black lines Acoustic lens damage Cable damage Connector damage Oil leakage If your instrument has any of the above problems or others, please feel free to
Kwaliteit Careutech BK Medical 9048 Ultrasone Probe Voor Reparatie Fabriek

Careutech BK Medical 9048 Ultrasone Probe Voor Reparatie

BK Medical 9048 Ultrasound Probe For Repair ► Brief Description Fault: The housing break. Solution: Replace probe housings. ►Problem & Solutoin Brand BK Medical Model 9048 Fault The housing break. Solution Replace probe housings. ► Common issues with 9048 Probes No Image / Poor Image Quality Probe Not Recognized by System Intermittent Connection Physical Damage Overheating If your instrument has any of the above problems or others, please feel free to contact us for
Kwaliteit Medische apparatuur onderdelen Sonoscape L743 lineaire probe voor reparatie Fabriek

Medische apparatuur onderdelen Sonoscape L743 lineaire probe voor reparatie

Sonoscape L743 linear probe for repair ► Brief Description Fault: Probe head broken. Solution: Replace probe head.(including crystal and upper housing). ►Problem & Solutoin Brand Sonoscape Model L743 Fault Probe head broken Solution Replace probe head.(including crystal and upper housing) ► Common issues with L743 Probes No Image or Intermittent Image Poor Image Quality Probe Not Recognized/Detected Physical Damage to the Probe Head or Cable Overheating Probe Gel Leakage If
Kwaliteit BK Medical 9096 Robotic Ultrasound Probe voor reparatie Fabriek

BK Medical 9096 Robotic Ultrasound Probe voor reparatie

BK Medical 9096 Robotic Ultrasound Probe For Repair ► Brief Description Fault: All the outer skin of the cable is seriously damaged. Solution: Replace the outer skin of the cable. ►Problem & Solutoin Brand BK Medical Model 9096 Fault All the outer skin of the cable is seriously damaged Solution Replace the outer skin of the cable ► Common issues with 9096 Probes Cable Damage Acoustic Lens/Membrane Damage Probe Housing Damage Crystal Array Damage Robotic Grasping Interface
Kwaliteit Samsung Medison 3D4-9 Probe SN P0N7 Voor reparatie Fabriek

Samsung Medison 3D4-9 Probe SN P0N7 Voor reparatie

Samsung Medison 3D4-9 Probe SN P0N7 For Repair ► Brief Description Fault: Scanhead broken. Solution: Replace scanhead ► Problem & Solutoin Brand Samsung Model 3D4-9 Fault Scanhead broken Solution Replace scanhead ► Common i ssues with 3D4-9 Probes Cable damage Lens delamination Element failure (3D/4D imaging issues) Connector problems If your instrument has any of the above problems or others, please feel free to contact us for assistance. We will do our best to help you
Kwaliteit Reparatie van medische hulpmiddelen Fabriek

Reparatie van medische hulpmiddelen

GE RIC5-9-D Probe Fault: Dome loose, straight relief tiny broken, with error on machine. Solution: Replace dome and oiling. Repair cost : $***. Common Issues with the GE RIC5-9-D Ultrasound Probe: Image Artifacts/Dropouts: These can manifest as lines, shadowing, or a complete loss of image in certain areas. Common causes include: Crystal Damage: Piezoelectric crystals within the transducer array can become damaged due to physical impact, overheating, or degradation over time.
Kwaliteit Hoekdraad en helixbuis Versleten CF-HQ190L 20% Gebroken lichtbundel Fabriek

Hoekdraad en helixbuis Versleten CF-HQ190L 20% Gebroken lichtbundel

Products: CF-HQ190L For Repair Problem & Solutoin: CF-HQ190L - Light bundle 20% Broken Knob Leaked The angle wire and helix tube Worn Biospy channel Worn Rubber Cracked Light guide tube Indented Solution Replace Biopsy Channel, Wire, Light Guide Tube, Knob, Rubber Applications: When it comes to medical device repair cases, the OLYMPUS CF-HQ190L is a reliable choice for healthcare professionals. This model is especially suitable for scenarios where the biospy channel is worn
Kwaliteit Reparatie van medische hulpmiddelen Olympus CF-H190L Colonoscoop voor reparatie Fabriek

Reparatie van medische hulpmiddelen Olympus CF-H190L Colonoscoop voor reparatie

Olympus CF-H190L Colonoscope for repair,SN:261****. Fault:Light bundle broken,biopsy tube worn,water air tube leakage,bending section deformation,wire helix tube worn. Solution:Replace the bending section,wire,helix tube,air tube and rubber.PS:Warranty for replace parts only. Repair time:7 working days. Repair price:$****. Warranty:90 days. Guangzhou Care techonoly Co., Ltd. is a company dedicated to the supply of medical equipment maintenance tools, maintenance accessories,
Kwaliteit Gerepareerde Olympus medische apparatuur Ultieme oplossing voor medische diensten Fabriek

Gerepareerde Olympus medische apparatuur Ultieme oplossing voor medische diensten

Product Description: Product Summary: The Medical Device Repair Cases product offered is a comprehensive solution for repairing and maintaining Olympus medical devices. With a warranty of 90 days, customers can have peace of mind knowing that their repaired devices are covered for an extended period. These repaired devices are in good condition, ensuring that they function effectively and reliably. The faults that have been addressed include light bundle broken, insert tube
Kwaliteit OLYMPUS CF-HQ190L Hersteldozen voor medische hulpmiddelen Fabriek

OLYMPUS CF-HQ190L Hersteldozen voor medische hulpmiddelen

OLYMPUS CF-HQ190L Repair Problem: 1, Image display;2, Light bundle broken about 10%;3, Insertion tube is ok;4, Biospy channel worn and leaking;5, The upper angle broken, helix tube and bending section were worn;6, Light guide tube indented and the back cover is leaking;7, The EL-connector is not good sealed and leaking. Solution: Replace biopsy channel, bending section, wire, helix tube, light guide tube, rubber Company Profile Guangzhou Careutech Co., Ltd. supplies medical