| Merk | Philips |
|---|---|
| Model | L15-7io |
| Schuld | Lensprobleem |
| Oplossing | Vervang de sondlens |
| Verpakking Details | Standaardpakket |
| Brand | OLYMPUS |
|---|---|
| Model | TJF-Q180V |
| Fault | 1. Image ok; 2. Light bundlestill ok; 3. Insert tubestill ok; 4. Biopsy tubewear; 5. Bendingsection, wires and helix tube wear; 6. Knob leakage. |
| Solution | Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD) |
| Verpakking Details | Standaardpakket |
| Brand | Olympus |
|---|---|
| Model | CF-240L |
| Fault | Leakage, glass of light bundle blur, rubber leak, angle not enough,image fuzzle,objective lens scratch, |
| Solution | Replace rubber, objective lens, C cover. Adjust the angle. |
| Packaging Details | Standard Package |
| Brand | Karl Storz |
|---|---|
| Model | 27005BA |
| Fault | 1. No image. 2. Fiber abrasion. 3. Rod lens broken and cracked. 4. Objective lens broken and dirty. 5. Eyepiece cover broken. |
| Solution | 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens group. 5. Clean the eyepiece. 6. Replace fiber. 7. Replace eyepiece cover. 8. Clean and install the optical path, check the image. 9. Seal with endoscope a |
| Packaging Details | Standard Package |
| Merken | Toshiba |
|---|---|
| Model | PST-25BT |
| Fout | Huisvestingsbreuk, beeld zwak, kristal veroudering |
| Oplossing | Vervang kristal en behuizing |
| Verpakking Details | Standaardpakket |
| Merken | Stryker |
|---|---|
| Model | 1188-210-105 |
| Fout | Afbeelding met onderbreking. |
| Oplossing | Vervang compatibele nieuwe kabel en connector |
| Verpakking Details | Standaardpakket |
| Merken | Storz |
|---|---|
| Model | H3-ZA |
| Fout | Beeld strepen vaak, kabel los contact en verouderd |
| Oplossing | Vervang de kabelkit |
| Verpakking Details | Standaardpakket |
| Merk | Ge |
|---|---|
| Model | RIC6-12-D |
| Schuld | Lensprobleem |
| Oplossing | Vervang lens, schone kabel |
| Verpakking Details | Standaardpakket |
| Merk | Hitachi |
|---|---|
| Model | C41V1 |
| Schuld | Lensprobleem |
| Oplossing | Vervang lens |
| Verpakking Details | Standaardpakket |
| Merk | Medtronic |
|---|---|
| model | 1898200T |
| Schuld | 1) Het lager is gecorrodeerd en de afdichting is kapot. |
| Oplossing | Vervang het hele handstuk, behalve de kabel. |
| Verpakking Details | Standaardpakket |